Determination of morpholine by gas chromatography

October 8, 2022by admin0
Morpholine is an important intermediate of fine chemistry and occupies an important position in chemical production. 60% of them are used as raw materials of rubber auxiliaries (sulfur promoter and antioxidant), and in medicine, it is used to synthesize many important drugs such as virus spirit, ibuprofen and cough. In addition, morpholine is also used as rust inhibitor, tank cleaner, neutralizer, surfactant, etc.
The traditional industrial production method of morpholine is based on the cyclization of diethanolamine by sulfuric acid dehydration. At present, the synthesis of morpholine by diethylene glycol and ammonia-catalyzed cyclization is the main production method at home and abroad. Analytical monitoring of the reaction mixture in production is an important part of controlling production. Regarding the analysis of morpholine samples, the determination of trace amounts of morpholine in boiler water has been reported abroad, but not in China. Since morpholine and its reaction mixture are somewhat alkaline, it is difficult to analyze them on a normal packed column. In this paper, a mixture of alkali-washed stretcher, SE-30 and PEG-20M stationary solution was used to quantify morpholine in the reaction mixture. The data showed that the results are accurate and reliable, and the method is suitable for production monitoring.
The GC 6891N laboratory gas chromatograph is based on an international advanced technology platform with internationally procured reliable components and hardware integration and specially designed stable software. High-quality hardware and intelligent operation and data processing software meet its long-term working measurement repeatability and convenience. The manufacturing precision, technical performance, and application flexibility are in the position of the industry.
Application areas
Petroleum extraction and refining, petrochemical, fine chemical, food safety, environmental protection, electronic high-purity gas industry, scientific research institutes, quality inspection, safety supervision, publican forensic and nationalan security and other fields of analysis and testing.
Chromatographic conditions
Column temperature
175℃
Vapor chamber temperature
270℃
Detector temperature
270℃
Carrier gas
N2
Gauge pressure
5.2×104Pa; hydrogen gauge pressure: 4.0×104Pa; air gauge pressure: 9.8×104Pa
Recorder
5mv, 4mm min-1